Journal of Ilam University of Medical Sciences 2018, 26(3): 72-81 Back to browse issues page
Calculation of peak to valley dose ratio due to multi-slice X-ray beams using ionization chamber and Geant4 simulation code
Mohammad reza Ghasemi *
Physics and Accelerators Research School, Nuclear Sciences and Technology Research Institute, Karaj, Iran , mghasemi842@gmail.com
Abstract:   (3567 Views)

  • Studies carried out with synchrotron radiation have shown that micro-beam radiation therapy (MRT) has unique advantages in the treatment of cancerous tumors. In this method, the determination of dose distribution and calculation of peak to valley dose ratio (PVDR) are considered as the most important steps in treatment planning. The PVDR is a criterion to evaluate the destruction of cancer cells and protection of normal cells in the tissues surrounding a tumor.

 
Materials and Methods: Using a multi-slit collimator, planar sliced beams were generated in an X-ray generator in order to determine dose distribution in a multilayer phantom made of plexiglass. An ionization chamber was used to measure absorbed dose. Given the large size of the sensitive area of the chamber in comparison with the narrow beams, a mono-slit collimator made of tungsten with a slit of 0.3×7.5 mm2 in its center was placed in front of the ionization chamber. Furthermore, by using Geant4 computer code, a model, including X-ray source, multi-slit collimator, phantom, mono-slit collimator, and detector, was designed to compare experimental and simulation results.
 
Findings: The investigation of dose distribution in the phantom with both methods indicated the presence of peaks and valleys. Given the low intensity of X-ray beam generated by the X-ray generator, and limited exposure time, the experimental errors were considerable. When using 1 mm (Air)+0.5 mm (W) collimator, PVDRs were obtained as 8.7 and 10.5 for ionization chamber and simulation, respectively, in the depth of 8 mm of the phantom. On the other hand, with a 1 mm (Air)+1 mm (W)  collimator, the values obtained for this parameter were 11.1 and 13.3 for ionization chamber and simulation, respectively.
 
Conclusions: Based on the results, a multi-slit collimator made of tungsten could produce multi-slice X-ray. The estimated dose distribution using the Geant4 code was more accurate than the one obtained through ionization chamber, which can be due to the possibility of using a detector in much smaller dimensions in the Geant4 code.

Keywords: Multi-sliced X-ray, Peak to valley dose ratio, Ionization chamber, Multi-slit collimator, Geant4 code
Full-Text [PDF 921 kb]   (1455 Downloads)    
Type of Study: Research |
Received: 2016/12/4 | Accepted: 2017/04/8 | Published: 2018/09/15
Send email to the article author

Add your comments about this article
Your username or Email:

CAPTCHA



XML   Persian Abstract   Print


Download citation:
BibTeX | RIS | EndNote | Medlars | ProCite | Reference Manager | RefWorks
Send citation to:

Ghasemi M R. Calculation of peak to valley dose ratio due to multi-slice X-ray beams using ionization chamber and Geant4 simulation code . J. Ilam Uni. Med. Sci. 2018; 26 (3) :72-81
URL: http://sjimu.medilam.ac.ir/article-1-3983-en.html


Rights and permissions
Creative Commons License This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.
Volume 26, Issue 3 (9-2018) Back to browse issues page
مجله دانشگاه علوم پزشکی ایلام Journal of Ilam University of Medical Sciences
Persian site map - English site map - Created in 0.15 seconds with 40 queries by YEKTAWEB 4646