Journal of Ilam University of Medical Sciences 2018, 25(5): 49-54 Back to browse issues page
Determination of PSF and SP from PDD for Mega Voltage Photon Beam
MohammadJavad TahmasebiBirgani1 , Ahlam Almanie * 2, MohammadAli Behroz1
1- Dept of Medical Physics, Faculty of Medicine, Jundishapour University of Medical Sciences, Ahvaz, Iran
2- Dept of Medical Physics, Faculty of Medicine, Jundishapour University of Medical Sciences, Ahvaz, Iran , maalmani90@gmail.com
Abstract:   (4320 Views)

Introduction: In radiation therapy, to calculate the dose absorbed by the patient, parameters such as Percentage Depth Dose (PDD), Tissue-Air Ratio (TAR), and Peak Scatter Factor (PSF)   etc. were used. For high energy photon beams, measurement of the TAR is subjected to some difficulties and because of the contamination of photon with electron it is inaccurate. The goal of this study was to find a formula for TAR from PDD and then calculate other dosimetry parameters as PSF and SP.
 
Materials & methods:  PDD at different treatment depths for field size 5 × 5 to 40× 40cm 2 was measured, then measuring the TAR for all treatment fields at the reference depth (10cm) and using the Tbl Curve Software, an analytical equation for TAR in 6MV and 18MV photon energies was obtained. Other treatment parameters S C , were calculated from this equation.
 
Findings: The PSF value was greater than one; this value ranged from 1.03 to 1.18 for 6MV and ranged from 1.01 to 1.10 for 18MV. Both PSF and SP increase with field, but didn’t have a significant and regular change with energy.
 
Discussion & Conclusion: There is a good consistency between obtained values and reported values. Determine of PDD in megavoltage treatment make calculate of TAR, PSF and SP easy and possible. Using this formula prevents repetition of measurements and also it prevents the force on the accelerator.
 

Keywords: Ionization chamber, Accelerator, Phantom scatter factor, Peak scatter factor
Full-Text [PDF 380 kb]   (1248 Downloads)    
Type of Study: Research |
Received: 2016/03/6 | Accepted: 2016/12/25 | Published: 2018/01/7
Send email to the article author

Add your comments about this article
Your username or Email:

CAPTCHA



XML   Persian Abstract   Print


Download citation:
BibTeX | RIS | EndNote | Medlars | ProCite | Reference Manager | RefWorks
Send citation to:

TahmasebiBirgani M, Almanie A, Behroz M. Determination of PSF and SP from PDD for Mega Voltage Photon Beam . J. Ilam Uni. Med. Sci. 2018; 25 (5) :49-54
URL: http://sjimu.medilam.ac.ir/article-1-3433-en.html


Rights and permissions
Creative Commons License This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.
Volume 25, Issue 5 (1-2018) Back to browse issues page
مجله دانشگاه علوم پزشکی ایلام Journal of Ilam University of Medical Sciences
Persian site map - English site map - Created in 0.16 seconds with 40 queries by YEKTAWEB 4646